Respiratory Analysis & Simulation

Adult/Infant Training and Test Lung

Versatility is power in the medical industry, and our model 1601 and 5601i test lungs feature both adult and infant lung chambers, giving them the widest possible range of applications.

  • Simulates a wide variety of pulmonary conditions. Lung compliance and resistances can easily be adjusted to simulate any type of pathology in either adult or infant scenarios. A training and test lung is a useful tool spanning respiratory therapy classrooms to hospital biomed departments.

  • Offers a wide range of compliance and resistance settings from 5-500 cmH­­2O/L/sec (resistance) and 0.001-0.10 L/cmH2O (compliance).

 

Model 1601 (analog) and 5601i (instrumented) units are currently available. Contact us for more information or to request a quote today!

 

  Download A Brochure  Request a Quote 

 

Parameter: Measurement Range: Accuracy:
Breath Rate: 2 to 200 bpm +/- 1 bpm
Inspiratory/Expiratoy Time: 0.3 to 30 sec +/- 0.03 sec
Tidal Volume:  (Infant lung)
.005 to 200 mL +/- 5% or 5 mL
Tidal Volume:  (Adult lung)
.100 to 2 L +/- 3% or 20 mL
Flow Rate:  (Infant lung) 3 to 80 Lpm (3 to 30 Lpm) +/- 2 Lpm
(30 to 80 Lpm) +/- 4 Lpm
Flow Rate:  (Adult lung) 10 to 200 Lpm

(10 to 60 Lpm) +/- 3 Lpm
(60 to 200 Lpm) +/- 5 Lpm

Pressure:  (Lung/Airway) -20 to 120 cmH2O +/- 0.5% of user-calibrated
Full Scale Output (FSO)
Pressure:  (Auxiliary Port-
Model 5601i only)
0 to 100 psi +/- 0.5% of user-calibrated FSO
Compliance:  (Infant lung) .001 to .010 L/cmH2O  
Compliance:  (Adult lung) 0.01 to 0.10 L/cmH2O  
Resistance: Rp5, Rp20, Rp50, Rp200 & Rp500
cmH2O/L/sec
 